Production Overview for concurrent batches: Incorrect alarm status color and alarm plots show additional batches
Symptoms
- Production Overview shows incorrect alarm status color for phase iteration, model or batch
- When clicking on a box in Production Overview for a specific batch, the plots created show other batches as well.
For this to happen, the following must be true:
- a batch project configuration with alarms is opened in the client
- there is an active batch
- the project configuration has overlapping batches and those batches have triggered alarms
Cause
The alarm status is confused between the active batch and any overlapping concurrent batches at the time when the project configuration was opened in the client.
More information
The problem is strictly client-side. It does not affect the server including alarm notifications.
Status
A fix will be made available in SIMCA-online 17.0.1.