Minisart® NY25 Plus Syringe Filter 1784B----------K, GF + 0.2 µm Polyamide (Nylon)
Minisart® NY Plus with glass fibre prefilter and hydrophilic polyamide (nylon) is optimized for aqueous liquids and solvents with a pH of 3 - 14. Type 1784B with a pore size of 0.2 µm can be used for sample preparation and sterile filtration. It has a male luer slip outlet. Version K is a non-sterile pack of 50. Units can be sterilized by autoclaving or EO.
Order Number: 1784B----------K
Availability: On Request
Configurate your Minisart® With Glass Fibre Prefilter
Non-sterile Minisart® NY Plus syringe filters are supplied in a pouch in one pack. This reduces waste and makes the handling easier whenever the application allows clarification under non-sterile conditions, like this is the case for the majority of sample preparation steps prior to HPLC or GC analysis. NY PLus is designed for highly particle-laden samples and combines leachable free GF prefilters and polyamide membranes with fastest flow and low non-specific binding. Sartorius polyamide is exceptionally pure compared with other common hydrophilic nylon filters. Competitor nylon membranes are using raw materials which are especially interfere with LC | MS. The housing of the filter is made of PP (polypropylene) which is compatibility with harsh solvents, acids and bases.
- A pouch in one pack for less waste
- Minimum extractables or leachables
- Low non-specific binding
- Outperforming chemical compatibility
Brochure
Catalog
Flyer
User Manual
Product properties
-
ColourWhite (opaque)
-
Filtration area4.8 cm²
-
Diameter25 mm
-
Flow rate50 mL/min with water at 1 bar (14.5 psi) | 70 mL/min with methanol at 1 bar (14.5 psi)
-
Pore size0.2 µm
-
Primary Product TypeFiltration Devices
-
BrandMinisart®
-
WettabilityHydrophilic
-
High FlowNo
-
Material prefiltersQuartz Fiber (QF)
-
PackagingBulk pouch packed
-
SterilityNon-Sterile
-
Connector OutletMale Luer Slip
-
Connector InletFemale Luer Lock
-
Membrane materialPolyamide (PA), Glass Fiber (GF) Prefilter
Technical attributes
-
Maximum Pressure4.5 bar
-
Operating Pressure4.5 bar (65 psi)
-
Hold-up volume230 µL before bubble point
-
Max. temperature121° C
-
Bubble pointMin. value with water 3.0 bar (44 psi)
-
Burst pressureMin. value 7.0 bar (102 psi)
-
Filter diameter25 mm
-
Housing dimensions33 × 24 mm
-
Device typeSyringe Filters
-
Sample TypeAqueous` Solvents` Harsh liquids according to chemical compatibility
-
Sterilization ProceduresEthylene oxide` Gamma irradiation` Autoclave, 121°C, 30 min
-
ExtractablesHPLC certified
-
Housing materialPolypropylene (PP)
-
Total throughputUp to 200 mL
-
ApplicationHPLC` Buffer & Media` Cell Removal / Harvest
-
Area of applicationSterile Filtration` Analytical Sample Preparation` Clarification
-
Sterilization MethodEthylene oxide` Gamma irradiation` Autoclave
-
CytotoxicityAccording to ISO 10993-5
-
pH Range3.0 – 14.0
No accessories & consumables found