Sartopore® 2 Gamma Midicaps® 0.2µm Size 0 | 2pc.

  • Effective Filtration Area: 0.45 m² | 4.8 ft²
  • Pore Size: 0.45 µm prefilter | 0.2 µm final filter
  • Connectors: Inlet: Hose Barb ½“ single stepped | Outlet: Hose Barb ½“ single stepped
  • >Sterilization: Gamma-irradiatable
  • Membrane Material: Polyethersulfone
  • Pack Size: 2 pieces
  • ________________________________________

    Connector explanation:

    O = Hose Barb ½“ | S = Sanitary 1 ½“ | H = Hose Barb ¼“ | F = Sanitary ¾“

    Size explanation:

    Size 7 = 0.065 m² | Size 8 = 0.13 m² | Size 9 = 0.26 m² | Size 0 = 0.52 m²

Price (USD):
$1,000.00
Item No.: 
5445307H0G-OO--V

*Custom/bulk order quotes are provided within 72 hours of request.

Sartopore® 2 Gamma Midicaps® were developed for use with bag containers prior to sterilization using gamma irradiation.

Applications

Typical applications include the sterile filtration and mycoplasma disposal of:

  • Biologicals
  • Pharma
  • Cell culture media (with and without serum)
  • Serum
  • Buffer
  • Culture media components

Compatibility

Sartopore® 2 Gamma Midicaps® have been developed for sterilization using gamma irradiation < in 50 kGy irradiation doses. The Sartopore® 2 polyethersulfone membrane has a high chemical compatibility between pH 1 to 14, making Sartopore® 2 Gamma Midicaps® ideally suited for the filtration of high and low pH buffers in pharmaceutical | Biotech Industry

Performance

Through "built-in prefiltration" in the form of a heterogeneous two-layer membrane, Sartopore® 2 Gamma Midicaps® achieve outstanding total flow and excellent flow rates.

Flexibility

Sartopore® 2 0.2 μm Midicaps® are supplied with different filtration areas of 500 cm² | 0.5 ft² to 0.45 m² | 5 ft² for easy adaptation to every filtration process regardless of batch size.

Microbiological Retention

Sartopore® 2 Gamma Midicaps® are completely validated as a sterile filter in accordance with HIMA ASTM F-838-05 guidelines. In addition, for mycoplasma retention, Sartopore® 2 Gamma Midicaps® are validated with an LRV of 7 for acholeplasma Laidlawii.

Quality

Each individual item was assigned before publication by a Bubblepoint and a diffusion test, integrity tested to ensure absolute reliability.

Documentation

Sartopore® 2 Gamma Midicaps® have been designed, developed and manufactured to comply with an ISO 9001 certified quality management system. A validation manual is available that meets regulatory requirements.

Applications

  • Application
    High capacity platform for conventional sterile filtration  
  • Industry
    Pharma / Biotech  

Biological Information

  • Bacterial Retention
    Quantitative Retention of 1 × E7 CFU/cm² Brevundimonas Diminuta ATCC® 19146 per ASTM methodology  

Compliance Information

  • Bacterial Endotoxins
    < 0.25 EU/mL as determined by the LAL test  
  • Biological Reactivity
    All materials of this filter element meet the requirements of the current USP Biological Reactivity tests <88> for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).  
  • Extractables
    The total amount of extractables is well below the limits established by the current USP unter "Sterile Water for Injection".  

Connections (Physical)

  • Connector Inlet
    1/2" Single Stepped Hose Barb  
  • Connector Outlet
    1/2" Single Stepped Hose Barb  
  • Connectors
    Hose barb 0.5" single stepped | Hose barb 0.5" single stepped  

Dimensions

  • Filter Size
    size 0 (0.45 m² / 4.8 ft²)  
  • Filtration Area
    0.45 m²  
  • Outer Diameter
    108.5 mm (77 mm without Connection / Valves)  

General Specifications

  • Capsule/Cartridge Format
    Midicaps®  
  • Maximum Allowable Back Pressure
    2 bar (29 psi) at 20°C; 1.5 bar (21.8 psi) at 50°C  
  • Maximum Diffusion
    at 20 °C and 2.5 bar | 36 psi: 0.05 m² | 0.5 ft²: 4 mL/min ; 0.1 m² | 1 ft²: 5 mL/min; 0.2 m²| 2 ft²: 7 mL/min; 0.45 m² | 4.5 ft²: 14 mL/min  
  • Minimum Required Bubble Point
    3.2 bar | 46 psi  
  • Vent Valve
    Standard  

Materials of Construction

  • Housing Material
    Polypropylene (PP)  
  • Material o-ring | seal
    Silicone (SI)  
  • Support Fleece
    Polyester (PET)  

Operating Conditions

  • Maximum Differential Pressure
    20 °C: 5 bar, 50 °C: 3 bar  
  • Maximum Operating Pressure
    5 bar (72.5 psi)  

Physicochemical Information

  • Non-Fiber Releasing
    This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.  
  • Pore Size Final-filter
    0.2 µm  
  • Pore Size Pre-filter
    0.45 µm  

Product Information

  • Brand
    Sartopore® 2  
  • Pack Size
    2  
  • Type
    Size 0 | Connectors: OO  

Sterility/Purity

  • Sterilization Method
    Gamma Irradiatable  

Usage Protocols

  • Sterilization Procedures
    Gamma Irradiation: with 50 kGy, 1 Cycle  
  • Sartocheck® 5 Plus Filter Tester
    Item no.
    26787---FT---P

    The Sartocheck® 5 Plus Filter Tester is an automatic filter integrity tester with automated Quality Risk Management (QRM) designed to cover the complete range of integrity testing from small syringe filters up to large multi-round housings and tangential flow cassettes.

  • Sartoguard
    Item no.
    M_Sartoguard

    All members of the Sartoguard family utilize a highly asymmetric polyether sulfone (PES) membrane double layer for unmatched protection of final filters and bioburden reduction with an LRV of 6 per cm². In addition, combinations with different depth filter materials and gamma sterilizable versions are available.