Skip to content Skip to navigation menu

CIMac™ Adeno 0.1 mL Analytical Column (2 µm)

CIMac™ Analytical Columns combine all elaborate characteristics of CIM monoliths at analytical scale and are your perfect tool for analysis and control of manufacturing processes, ideal for process analytical technology (PAT). Designed for fast and reproducible HPLC monitoring and quantitation of proteins, viruses, pDNA, and other large biomolecules, they allow you to achieve rapid, high-resolution and flow-independent separations in a matter of minutes.

Item no.:  110.8502-2

   

The CIMac™ Adeno 0.1 mL Analytical Columns are intended for fast and reproducible HPLC monitoring and quantitation of any type of Adenovirus.

Adeno features:

  • Strong anion exchanger
  • Enables fast and reproducible HPLC monitoring and quantitation of any type of Adenovirus

 

  • Channel Size
    2 µm
  • Connectors
    10-32 UNF coned port, 1/16’’ OD tubing connection
  • Diameter
    5.2 mm
  • Recommended Operating Temperature
    4 °C (39 °F) to 40 °C (104 °F)

 

  • Bed Volume
    0.1 mL
  • Chemical Compatibility
    All commonly used aqueous buffers, 1 M NaOH, 0.1 M HCl, 8 M urea, 6 M guanidine hydrochloride and 20 % ethanol solution. Avoid oxidizing agents. Avoid prolonged use of concentrated acids (more than 0.5 M) like hydrochloric, sulphuric or acetic acid.
  • Flow Rate
    0.2 – 3 mL/min | 1 – 15 cm/min | 2 – 30 CV/min
  • Ligand
    QA modified for Adeno (strong anion exchanger, quaternary amine)
  • Materials of Construction
    Poly(glycidyl methacrylate -co- ethylene dimethacrylate)
  • Maximum Operating Pressure
    15 MPa | 150 bar | 2175 psi
  • Reusable
    Yes
  • pH Stability
    Working range 2–13, cleaning in place 1–14

 

  • Application
    Analytical Columns
  • Length