Skip to content Skip to navigation menu

Sartocheck® 5 Plus Filter Tester

The Sartocheck® 5 Plus Filter Tester is an automatic filter integrity tester with automated Quality Risk Management (QRM) designed to cover the complete range of integrity testing from small syringe filters up to large multi-round housings and tangential flow cassettes.

Item no.:  26787---FT---P

   

The Sartocheck® 5 Plus Filter Tester is an automatic filter integrity tester designed to cover the complete range of integrity testing from small syringe filters up to large multiround housings and tangential flow cassettes.

The Sartocheck® 5 Plus can perform the following filter integrity tests:

  • Bubble Point Test
  • Diffusion Test / Forward flow Test
  • Combined Diffusion and Bubble Point Test
  • Water Intrusion Test (WIT)
  • Pressure Drop Test / Leak Test

Quality Risk Management

The regulatory focus on QRM (cf. ICHQ9 and the new Annex 1 written by EMA in cooperation with the US-FDA, WHO and PICs) also applies to filter integrity testing, as a fundamental element of sterility assurance. The Sartocheck® 5 Plus Filter Tester uses program specific parameters allowing the automatic identification of testing anomalies before or during the test. This prevents time-consuming, costly deviations, potential drug recalls and 483 warning letters.

Main Features:

Quality Risk Management

  • Automatic detection of incorrect test setups
  • Automatic detection of abnormal test conditions
  • Self-test at booting and before each test

Data Integrity

  • Custom Linux-based Operating System
  • Four eyes principle with electronic signature
  • C10- Internal Double Data Backup
  • Encrypted Data
  • Time Zone synchronization - NTP
  • Audit Trail with time stamped events
  • User Matrix to generate individual user roles
  • Mandatory fields for Operator entries
  • Data entry by Barcode Scanner
  • Long-life printouts and sticker function from optional printer

Usability

  • Intuitive iF-Design rewarded Human Machine Interface (HMI)
  • 1"" bright touchscreen with a +-88° viewing angle
  • Large digital keypad - no need for a pen – compatible with glove use
  • 10 system languages
  • LDAP - Log on with network user credentials
  • Automatic test time for faster testing
  • Data transfer

Health, Safety and Environment

  • Splash-proof (IP64)
  • Explosion proof (ATEX IECEx & FM certified
  • Safe testing and re-testing of alcohol wetted filters
  • Continuous and clear visualization of pressure status
  • Resistant to all current cleaning agents
  • H2O2-vapor resistant (VHP)
  • PFA tubes (FDA 21 CFR 177 and USP Class VI compliant)
  • Prepared for automated cleaning of all pneumatics with up to 0.5 M NaOH

Optional accessory kit for backflow protection Compliance

  • 21 CFR part 11
  • GAMP 5
  • Ex proof (ATEX | IECEx | FM)
  • CE
  • FCC

Supplies

  • Sartocheck® 5 Plus Filter Tester
  • Inlet and outlet tubes
  • Country specific power cable
  • Instruction manual
  • QRM documentation - FMEA
  • Network cable 2m
  • USB- Cable for the optional printer
  • T20 screw driver

Datasheet

Sartocheck® 5 Plus Filter Tester
en 993.0 kB

 

  • Bubble Point
    250 – 6,550 mbar acc. +/- 50 mbar
  • Data Interface
    • USB
    • RJ45
  • Diffusion
    Without ext. ref. tank: Cartridge systems up to net volume of 14 l; with ext. ref. tank: Cartridge systems up to 150 l
  • Languages
    English, German, French, Spanish, Italian, Mandarin (simplified), Portuguese (Brazilian), Japanese, Korean, Russian
  • Maximum Operating Pressure
    8 bar (116 psi)
  • Measuring Accuracy - Absolute Deviation Pressure Measurement
    ± 7.2 mbar
  • Measuring Accuracy - Relative Deviation Pressure Measurement
    ± 0.1% full scale
  • Minimum Inlet Pressure
    4,000 mbar
  • Net Volume Measurement
    4 %
  • Operating Conditions - Temperature
    0 °C – 40 °C (32 – 104 °F)
  • Power Requirements
    100 – 240 V AC at 50 | 60 Hz
  • Pressure Drop
    All filter systems up to net volume of 150 L
  • Pressure Measuring Range
    50 – 6,600 mbar
  • Relative Humidity
    10% – 80%
  • System Type
    Integrity Testing Systems
  • Water Intrusion
    0.005 - 300.000 ml/min

 

  • Input Wattage
    74 W
  • Market
    BioPharma
  • Operating Pressure
    min. 4 bar g or test p + 0.5 bar g (whichever highest); max 8 bar g

 

  • Additional Product Specifications
    • Measuring accuracy - bubble point: 250-6
    • 550 mbar acc. +/-50 mbar
    • Measuring accuracy - diffusion: 0.01-24.000 mL/min acc. +/-5%
    • Measuring accuracy - Water intrusion: 0.005-600.000 mL/min acc. +/-5%
    • Measuring Ranges - Net volume: 10 mL - 150 L acc. +/- 4%
Self-adhesiv paper+thermotransf.foil 90m

Printingpaper + thermotransferfoil 90m

Thermopaper 5 x 24m
Thermopaper 5 x 24m
Item no.: 69Y03287