text.skipToContent text.skipToNavigation

Minisart® PES15 Syringe Filter 1776D--------ACK, 0.2 µm Polyethersulfone

Minisart® PES with polyethersulfone (PES) is optimal for highest flow rates and a pH of 2 - 13. Type 1776D with a pore size of 0.22 µm can be used for sterile filtration and has a male luer slip outlet. Version ACK is individually packed and sterilized by ethylene oxide (EO).

Order Number:  1776D--------ACK

Price:  $145.00

Availability:  In Stock *

*Detailed delivery date provided during check-out

Individually packed and sterile Minisart® PES syringe filters are optimal for immediate usage whenever sterile filtration or clarification and cell removal under sterile conditions is needed. Due to the slightly asymmetric membrane structure, the PES surface almost behaves like a pre-filter and enables highest flow rates. The low binding membrane is tightly connected with the housing which allows bidirectional use. The housing is made of PP (polypropylene) which is compatibility with harsh solvents, acids and bases.

  • Sterile and ready-to-use
  • Slightly asymmetric PES with superior flow rate
  • Low non-specific binding
  • Bidirectional use possible

Brochure

Minisart® Syringe Filters — The Easy Choice – Clean and Safe (US and Canada)
en 6.4 MB
Шприцевые насадки Minisart® — Лёгкий выбор, высокая степень чистоты и надёжности
ru 2.7 MB

Catalog

Laboratory Filtration Products (US and Canada)
en 9.6 MB

User Manual

Minisart® Chemical Compatibility Guide
en 72.6 kB
HPLC-Certificate — Minisart® Syringe Filters for Sample Preparation – reliable and ultrapure
en 364.1 kB
Minisart®-RC | SRP | NY | PES — Spritzenvorsatzfilter von Sartorius, verlässlich und hochrein
en 1.2 MB

Product properties

  • Colour
    White (opaque)
  • Filtration area
    1.7 cm²
  • Diameter
    15 mm
  • Flow rate
    40 mL/min with water at 1 bar (14.5 psi)
  • Pore size
    0.2 µm
  • Primary Product Type
    Filtration Devices
  • Brand
    Minisart®
  • Wettability
    Hydrophilic
  • High Flow
    No
  • Packaging
    Single blister packed
  • Membrane material
    Polyethersulfone (PES)
  • Connector Outlet
    Male Luer Slip
  • Connector Inlet
    Female Luer Lock
  • Sterility
    Sterile

Technical attributes

  • Operating Pressure
    4.5 bar (65 psi)
  • Hold-up volume
    30 – 100 µL before bubble point
  • Max. temperature
    121° C
  • Bubble point
    Min. value with water 3.2 bar (46 psi)
  • Burst pressure
    Min. value 7.0 bar (102 psi)
  • Filter diameter
    15 mm
  • Housing dimensions
    25 × 23 mm
  • Device type
    Syringe Filters
  • Sample Type
    Aqueous` Solvents and harsh liquids according to chemical compatibility
  • Biosafety
    All materials of this filter element meet the requirements of the current USP Biological Reactivity tests for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
  • Non-Fiber Releasing
    This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
  • Housing material
    Polypropylene (PP)
  • Total throughput
    Up to 15 mL
  • Application
    Buffer & Media
  • Area of application
    Sterile Filtration` Clarification
  • pH Range
    3.0 – 14.0
  • Sterilization Method
    Ethylene oxide
  • Cytotoxicity
    According to ISO 10993-5

No accessories & consumables found