Skip to content Skip to navigation menu

Minisart® Syringe Filter, Polytetrafluorethylene (PTFE), Pore Size 0.2 µm, Ethylene Oxide, Female Luer Lock, Needle (38 × 1.7 mm), Pack Size 50

Minisart® Air with chemically inert hydrophobic PTFE is suitable for venting as well as filtration of gases and oils with a pH of 3 - 12. Type 16596HNK with a pore size of 0.2 µm can be used for sterile filtration. It has an outlet with needle. Version HNK is individually packed and sterilized by ethylene oxide (EO).

Item no.:  16596--------HNK

Price:  $230.00

Availability (ex works):  For more information please log in

Individually packed and sterile Minisart® Air syringe filters are optimal for immediate usage whenever sterile filtration or venting is needed. The low-binding PTFE membrane is the best option for leachable free filtration. Due to its hydrophobic nature the PTFE membrane functions as a water barrier and does not become wetted by aerosols. This keeps the hydrophobic venting filter permeable for gas or air! The housing of the filter is made of MBS (Methacrylate butadiene styrene) which has a limited compatibility with alcohols but is free of phthalates and passes the plastic class VI test. Housing and membrane are tightly connected which allows bidirectional use.

  • Sterile and ready-to-use
  • Minimum extractables or leachables
  • Very low non-specific binding
  • Suitable as venting filter

Brochure

Шприцевые насадки Minisart® — Лёгкий выбор, высокая степень чистоты и надёжности
ru 2.7 MB
宣传手册:微生物质量控制产品精简目录
zh 2.4 MB
宣传手册:实验室过滤产品精简目录
zh 19.0 MB

Catalog

Laboratory Filtration Products (US and Canada)
en 9.6 MB

Datasheet

Venting Needle for Ampoules, Collapsible Bags and Vials
en 68.5 kB

User Manual

Minisart® Chemical Compatibility Guide
en 72.6 kB
Minisart® High Flow | NML | Air | HY | Acticosart — Sartorius Spritzenvorsatzfilter für die Filtration von Flüssigkeiten und zur Belüftung
en 1.2 MB

Applications

  • Application
    Buffer & Media
  • Area of Application
    • Sterile Filtration
    • Venting / Moisture Barrier

Compliance Information

  • Biosafety
    All materials of this filter element meet the requirements of the current USP Biological Reactivity tests for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
  • Cytotoxicity
    According to ISO 10993-5

Connections (Physical)

  • Connector Inlet
    Female Luer Lock
  • Connector Outlet
    Needle (38 × 1.7 mm)

Dimensions

  • Filter Diameter
    ⌀ 15 mm
  • Filtration Area
    1.7 cm²
  • Housing Dimensions
    25 × 72 mm

Features

  • High Flow
    No

Materials of Construction

  • Housing Material
    Acrylic-based multipolymer compound (MBS)
  • Membrane Material
    Polytetrafluorethylene (PTFE)

Operating Conditions

  • Maximum Allowable Working Temperature
    60 °C (140 °F)
  • Maximum Inlet Pressure
    3.0 bar (44 psi)
  • Maximum Operating Pressure
    4.5 bar (65 psi)
  • Operating Pressure
    4.5 bar (65 psi)

Physicochemical Information

  • Bubble Point
    Min. value with ethanol 0.9 bar (13 psi)
  • Burst Pressure
    Min. value 7.0 bar (102 psi)
  • Non-Fiber Releasing
    This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
  • Pore Size
    0.2 µm
  • Wettability
    Hydrophobic

Product Information

  • Device Type
    Syringe Filters
  • Pack Size
    50
  • Packaging
    Single blister packed
  • Primary Product Type
    Filtration Devices
  • Sample Type
    • Air/Gas
    • Oils and alkanes according to chemical compatibility

Specifications

  • Filter Paper/Membrane Color
    Yellow
  • Flow Rate
    2000 mL/min with air at 0.1 bar (1.5 psi)
  • Maximum Allowable Back Pressure
    0.3 bar (4.4 psi) at 20°C
  • pH Range
    3.0 – 12.0

Sterility

  • Delivery Condition
    Ethylene Oxide

Usage Protocols

  • Sterilization Method
    Ethylene oxide

Volumes

  • Hold-up Volume
    230 µL before bubble point