» O2 Enrichment
» Gas Flow Ratio Control
» Advanced Additive Flow 2 gas outlets
» Advanced Additive Flow 6 gas outlets
» Exclusive Flow

Exclusive Flow | |
The Exclusive Flow gassing strategy uses a series of 3-way solenoid valves to select either Air, O2, N2 or CO2 to flow the Sparger with Air as the default. Either Air or CO2 also flows to Overlay. O2, N2 and CO2 can be switched from automatic to manual control. Flow meters for Sparger and Overlay visually indicate the flow of the selected gas. The design does support the use of mass flow controllers for total flow to Sparger and Overlay. |
|